Volume Change Tool Kit for NEXTY Fixed Volume Pipettes

You can change the volume with the Volume Change Tool Kit and use it as your original volume pipette.

  • Optimize pipetting work of the samle volume in your daily routine, with the use of variable volume pipettes.
  • Reduce error in volume setting.
  • Reduce disposal costs of pipettes due to contamination.
  • Change the volume to the most suitable for your experiments.

Disassembly

Please note that the specifications of the fixed cap have changed starting from the summer 2014 manufacturing batch.

Please check which type of product you have, and if it is the old type, refer to section ⑤. If the fixed cap of the old type is difficult to open, please use a minus screwdriver or similar tool to pry it open.

Adjustment

1.Unlock
2.Dial adjustment
3.Lock
4.Measurement

Assembly

Please note that the new type has a tighter fit for the fixed cap, so please firmly press it all the way in with both hands. For details, please refer to the instruction manual and the video provided below.

※Please note that the new type has a tighter fit for the fixed cap, so please firmly press it all the way in with both hands. For details, please refer to the instruction manual and the video provided below.

Kit items

Lock-spanner:1pc.
Manual
Seal:12pcs.

Necessary tools to change volume

  • Electoric Balance(Prepare a balance with the readability according to the required precision.)
  • Pipette Tip(compatible to NEXTY pipette)
  • Vessel(placed on the balance to receive distilled water for measurement)
  • Distilled Water
  • Thermometer、hygrometer、barometer
Cat. No.ItemUnitBuy from Online Shop
(for the U.S.A.)
NT-FVCVolume Change Tool Kit1 unit

Product Inquiry

※We also accept inquiries for mass production development of custom shapes, such as OEM and ODM, for plastic products.